The discharge measurement is performed by using the proven tracer-dilution method and is especially suitable for high volumes of water. It can be deployed in turbulent rivers, creeks or waters for which data regarding the cross-section profile are not available. As a tracer a fluorescent material is used. After insertion of a known amount of tracer material two optical fluorescence probes are determining the discharge.

FEATURES AND ADVANTAGES

  • Simple, mobile discharge measurement
  • Reliable results even without knowledge and independent of the cross-section profile
  • Suitable for fast-flowing, turbulent waters or where access is limited
  • Applied especially with water amounts of several m³/s
  • Environmentally friendly, harmless for water
  • Discharge values immediately available
  • Easy transmission of measurement data form the sensor to receiving device via Bluetooth
  • Convenient receiving of data on laptop
  • Real-time visualisation of measurement data
  • User-friendly guidance through the measurement by the included software TQ-Commander
  • Simultaneous measurement with up to four probes
  • Tracer material: fluorescence (fluorescein or similar)
  • Probe type: optical fluorescence probe
  • Compact and convenient: complete equipment securely packed in two cases

TECHNICAL DETAILS TQ-F

GENERAL

  • Measuring principle tracer dilution method with instantaneous feed
  • Application all discharge volumes

TQ-AMP (MEASUREMENT DEVICE WITH BLUETOOTH-TRANSMISSION)

  • Memory capacity none (data storage in the receiving device)
  • Transmission interval 1 second
  • Data transfer Bluetooth class 1 (transmission range up to 100m)
  • Operating temperature -20 … +60 °C
  • Protection IP66
  • Energy supply 3x 1.5 V batteries, size AA or 3x 1.5 V / 2500 mAh NiMH batteries, size AA
  • Operation time 25 hrs (with 3x 2500 mAh batteries)
  • Recharging time approx. 10 hrs

PROBES

  • Probe type fluorescence probe
  • Measurement range Fluorescein/Rhodamine: 0 … 50 μg/l (ppb)
  • Resolution 0.05 μg/l (ppb)
  • Operating temperature 0 … +50 °C
  • (Further types of probes for different tracer substances available upon request.)

ACCESSORIES INCLUDED

  • Pipette 500 μl
  • Jars bottle for calibration solution, measuring cup stainless steel 750 ml, measuring cup 500 ml, volumetric flask 500 ml
  • Others USB memory stick (software and documentation), USB Bluetooth adapter, recharger, protection- and weighting armour